Drill Unit & Handpieces

Description
30,000 RPM Electric Drill Unit (110V)
 
30,000 RPM Electric Drill Unit (220V)
 
18:1 Contra-Angle Handpiece
 
400:1 Contra-Angle Handpiece
 

 

 

Description
20,000 RPM Air Motor
 
Contra-Angle Handpiece Head
Pana Spray*


*Available in USA only (ground shipping).

 

 

Handpiece Considerations

• The 400:1 handpiece has sufficient torque to cut through bone at 50 RPM.

• The 18:1 handpiece is designed to be used with the pilot drill at ~1,100 RPM with external irrigation.

• Irrigation is not necessary at speeds of 50 RPM or lower.

• Failure to observe speed limitation may result in burned or overheated bone which could result in substantial bone necrosis.

• The 20,000 RPM Air Motor attaches to conventional four hole dental tubing and is an alternative to the electric drill unit.

• Air pressure of at least 80 PSI is recommended to drive the air motor with sufficient torque in dense bone.

• Proper maintenance of handpieces is crucial for long term success of handpieces.

• Do not exceed 132°C (275°F) when sterilizing.
• In order to prevent discoloration and/or damage to the plating of the handpieces from chemicals that are not sufficiently cleaned from other instruments, do not autoclave the handpieces with other instruments.

• Always check the handpiece for any abnormal vibration, heating, noise, or sluggish operation. If any abnormality is noticed, cease the use of the handpiece.
 


 